The FM flowmeters are devices for the supply of medical gases with two to six antistatic and graduated tubes and completed with an integrated dosage unit. The body is made of anodized aluminum and the inlet connections for O2, Air and N2O at the bottom of the flowmeter are arranged to be connected to hospital centralized gas plants or to cylinders complete with pressure regulators. The gas feeding unit includes a safety valve automatically blocking the N2O supply when the feeding pressure of O2 drops below the value of approximately 0.8 bar. The mixtures of O2 and N2O can be supplied to the patient through a dosage unit. An incorporated mixing device proportions the N2O in such a way that a minimum of at least 30% of O2 is always present in the gas mixture. At the outlet of the dosage unit, an anesthesia evaporator can be connected or the gas mixture can be supplied directly to the patient through the gas feeding unit. The O2+ flush button can be used for a quick supply of O2 directly at the outlet of the mixture on the gas feeding unit.

FM 2200
FM 2300
FM 2500
Sizes (LxWxH)
160x145x318 mm
160x145x318 mm
160x145x318 mm
Weight
5.52 Kg
5.57 Kg
5.64 Kg
Dosage range
O2: 0.2 - 15 L/min.
N2O: 0.2 - 12L/min.
O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
N2O:
0.1 - 1 L/min.
0.2 - 12 L/min.
O2: 0.2 - 15 L/min.
Air: 0.2 - 15 L/min.
N2O: 0.2 - 12 L/min.
FM 2800
FM 2900
Sizes (LxWxH)
160x145x318 mm
160x190x318 mm
Weight
5.69 Kg
7.2 Kg
Dosage range
O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
Air:  0.2 - 15 L/min.
N2O::
0.1 - 1 L/min.
0.2 - 12 L/min.
O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
Air:  0.2 - 15 L/min.
N2O:
0.1 - 1 L/min.
0.2 - 12 L/min.
CO2 0.1 - 2 L/min. (limited at 1 L/min.)
Accuracy
±10% read value or ±0.3 L/min. if greater (± 10% E.O.S. for flow < than 1 L/min.)
Inlet pressures
3.5 - 5 bar ± 20%
N2O blocking device
the supply of N2O is blocked when the O2pressure drops below 0.8 bar ±0.3 bar
Charge loss level, side under pressure (before the adjusting needle valves)
less than 25 ml/min. in normal pressure conditions (ISO 5358)
Charge loss level, low pressure side(after the adjusting needle valves, vaporizer excluded)
less than 25 ml/min. at 30 cm H2O
Hypoxyguard system
at least 30% -3% O2 is garanteed in the mixture with N2O opening value equal to 0.4 L/min. (with inlet pressures of 3.5 bar)
Mixed gases outlet connection
conical connection 23 mm F ISO DIN 5356/1
Gases inlet connections
hose connector Ø 6 mm
O2 quick dosage
(if foreseen)
when pushed it can supply more than 45 L/min. O2